82 m system
Measurements
of the vertical flux of CO2 and H2O
by eddy covariance
82 m above the ground began in April 1997.
This was facilitated by mounting to the 4.4 m
long instrument arm an ultrasonic anemometer (Gill
Solent Enhanced), an aspirated thermocouple (50
cm from the anemometer) for fast-response temperature
measurements, and an additional air sampling tube
for CO2 and H2O measurements.
A schematic diagram
of the air sampling system used to determine the CO2 and
H2O mixing ratios for flux calculations.
Measurements of
CO2 and H2O are made at 4 Hz using a fast response
IRGA (Li-Cor model LI-6262). Air
is pumped (dual diaphragm GAST model DAA V175 ED MFG Corp.) through
the sampling tube and analyzer at about 15 l/min, producing a pressure
drop of approximately 45 kPa. Pressure fluctuations generated by the
pump are damped by means of a 6 l buffer volume. Flow in the sampling
tube is turbulent. Immediately behind the sample cell of the analyzer
we measure pressure (MKS Instruments model 122A barotron), temperature
and relative humidity (Vaisala HMD20YB). The pressure and temperature
data are used to correct the instrument response for variations in these
parameters (see below). The humidity data are used to determine the
calibration function for water vapor measurements by the Li-Cor 6262.
The CO2
analyzer runs in relative mode. Dry, synthetic air with a CO2
mixing ratio of 330-340 ppm is used as a reference gas (Messer Hungarogáz)
and flows at a rate of 5-10 cm3/min through the reference
cell of the IRGA. The analog output of the analyzer for CO2
and H2O, as well as the signals of the pressure and temperature/humidity
sensors are digitalized by the common A/D-RS232 converter.
A
separate data acquisition computer is used to
read data from the fast-response instruments (sonic
anemometer, thermocouple, and IRGA). The computer
communicates with the instruments through two
standard serial ports: COM1 receives the data
from the sonic anemometer while the two A/D-RS232
devices are controlled via COM2. The data acquisition
cycle is triggered by the signal from the sonic
anemometer. After reception of a data package
(horizontal and vertical wind speed, wind direction
and error code) the computer requests data from
the aspirated thermocouple, the CO2/H2O
analyzer and its accessory sensors (pressure,
temperature/humidity), and the CO2
profile analyzer. The position of the multiport
valve of the profile system is also determined.
As the time of the change of the position of the
multiport valve is recorded on the computers of
both the profile and eddy correlation systems,
the parallel records allow the synchronization
of the two independent data acquisition computers.
The data acquisition software is written in Turbo
Pascal language and runs under DOS 5.0. The eddy
correlation system produces data at a rate of
about 600 MByte/month, and the data are stored
on a CD-R without compression.
The response of
the IRGA is calibrated by comparison to ambient CO2 and H2O
measurements from the slow-response sensors. A calibrated CO2
measurement is typically obtained every 8 minutes at the 82 m level.
Exact synchronization of the signals requires accounting for the time
for air to pass through the sampling tubes of the profile and eddy flux
systems. The lag time of the profile system is calculated from the measured
flow rates. For the eddy correlation system two different lag time values
are calculated for H2O and CO2 using a spectral
method.
For calibration
of the H2O and CO2 response of the IRGA we use
the following function:
f(V*p0/p)*(T/T0)
+ Cr = C
where V is the voltage
signal of the analyzer, p0 is the reference pressure (1000
hPa), p is the pressure in the measuring cell, Cr is the mole fraction
of CO2 or H2O in the reference gas (zero for H2O,
330-340 mmol/mol for CO2), T0 is the reference
temperature (273.15 K), T is the temperature in the measuring cell,
and C is the mole fraction of H2O or CO2 measured
by the calibrated (profile) instruments. The analyzer is calibrated
in terms of H2O and CO2 mole fraction. Linear
regression is carried out between (C-Cr)*T0/T and V*p0/p
to determine the slope of the response function, f. Occasional outliers
are removed interactively, and the resulting linear fits typically show
very high correlation. The manufacturer provides a fifth order polynomial
calibration curve for the instrument, but the polynomial differs only
slightly from linear in the range of interest (360-500 ppm). Calibration
values are determined for each 24 hours of measurement. The calibration
factors are quite stable in time. Since the fluctuations of H2O
and CO2 measured by the Li-Cor 6262 are expressed in terms
of mole fraction relative to dry air it is not necessary to perform
the corrections for variations of the density of air.
Three dimensional
wind vector rotation is applied to the sonic anemometer data. Rotation
angles and measured average vertical wind speed are calculated and stored
for each hourly period.
A linear trend is
removed from each 60 min interval of all data used for eddy flux calculations
(wind components, temperature, H2O and CO2). Data
values outside ± 4st.dev. are removed. Turbulent fluxes are calculated
from covariances of the detrended time series taking into account the
delay time of signals determined during the calibration procedure outlined
above.
Spectral corrections
are applied in order to account for the damping of fluctuations caused
by the long air sample tubes, limited sensor response time, sensor line
averaging and sensor separation, and sampling. Average losses of the
eddy-correlation system calculated from the theoretical considerations
are about 10% for CO2 and 9% for H2O. It appears
that the water vapour signal suffers from excess spectral degradation
compared to the theoretical spectral damping because of the long tubing
of the system which causes extra loss of water vapour flux.
Net
ecosystem exchange (NEE, the sum of the eddy flux
at 82 m, the rate of change of CO2
storage below 82 m) is calculated from the eddy
correlation data using the profile data for storage
calculations using surface layer similarity theory.
The
NEE measurements are representative to a larger
region around the tall tower. The
tower is surrounded by agricultural fields (mostly
crops and fodder of annually changing types) and
forest patches. The distribution of vegetation
types (60% arable land, 30% forest and woodland,
10% other) within 10 km of the tower is not greatly
different from the average for the Western Hungarian
Landscape Unit or the whole country (85% of the
area is cultivated, 77% of which is agricultural
and 23% is forest).
3
m system
The
second eddy-covariance system owned by AIST
(the former NIRE, Japan) is operated in the garden
of the TV tower, above semi-natural grass (Barcza
et al., 2003). In the beginning the measuring
system was operational during the period of 1999-2000.
Due to malfunction of the anemometer the measurements
were ceased in 2001. In September 2006 the measurements
were resumed when the ultrasonic anemometer was
replaced with a new model (Solent Research R3-50,
Gill Instruments Ltd., Lymington, United Kingdom).
The
sonic anemometer and the inlet tube of the IRGA
are mounted on a mast at 3 m elevation above the
grass covered surface. The inlet tube is mounted
at the elevation of the active center of the ultrasonic
anemometer, 25 cm away from it horizontally. Raw
voltage data generated by the fast response sensors
were collected and digitized by means of a TEAC
data logger between 1999 and 2000, while since
2006 the digital signals of the sensors have been
recorded by a PC also at 5 Hz frequency.
The
area of the site was used as arable land previously
and was turned into a grassland around 1990. The
dominant species of the grassland are Arrhenatherum
elatius, Taraxacum officinale, Poa
pratensis, Agropyron repens, Anthoxanthum
odoratum, Dactylis glomerata, Holcus
lanatus, Briza media and Festuca
pratensis. The grass is mowed two times a
year, and the mowed grass is taken away from the
site and utilized as fodder. Fore up-to-date results
please see Nagy et al. (2010).
For
more details please see:
Nagy,
Z., Barcza, Z., Horváth, L., Balogh, J., Hagyó,
A., Káposztás, N., Grosz, B., Machon, A., Pintér,
K., 2010. Measurements and estimations of biosphere-atmosphere
exchange of greenhouse gases -- Grasslands. In:
Atmospheric Greenhouse Gases: The Hungarian Perspective
(Ed.: Haszpra, L.). Springer, Dordrecht - Heidelberg
- London - New York, pp. 91-119. ISBN 978-90-481-9949-5,
e-ISBN 978-90-481-9950-1, doi: 10.1007/978-90-481-9950-1
Tóth,
E., Barcza, Z., Birkás, M., Gelybó, Gy., Zsembeli,
J., Bottlik, L., Davis, K. J., Haszpra, L., Kern,
A., Kljun, N., Koós, S., Kovács, Gy., Stingli,
A., Farkas, Cs., 2010: Measurements and estimations
of biosphere-atmosphere exchange of greenhouse
gases -- Arable lands. In: Atmospheric Greenhouse
Gases: The Hungarian Perspective (Ed.: Haszpra,
L.). Springer, Dordrecht - Heidelberg - London
- New York, pp. 157-197. ISBN 978-90-481-9949-5,
e-ISBN 978-90-481-9950-1, doi: 10.1007/978-90-481-9950-1
Barcza,
Z., Kern, A., Haszpra, L., Kljun, N., 2009. Spatial
representativeness of tall tower eddy covariance
measurements using remote sensing and footprint
analysis. Agricultural and Forest Meteorology
149, 795-807. doi: 10.1016/j.agrformet.2008.10.021
Haszpra, L., Barcza, Z., Davis, K. J., Tarczay,
K., 2005. Long-term tall tower carbon dioxide
flux monitoring over an area of mixed vegetation.
Agricultural and Forest Meteorology,
132, 58-77.
Bakwin,
P. S., Davis, K. J.,Yi, C., Wofsy, S. C., Munger,
J. W., Haszpra, L., Barcza, Z., 2004. Regional
carbon dioxide fluxes from mixing ratio data.
Tellus B, 56 (4), 301-311.
Barcza, Z., Haszpra, L., Kondo, H., Saigusa, N.,
Yamamoto, S., Bartholy, J., 2003. Carbon exchange
of grass in Hungary. Tellus B, 55 (2),
187-196.
Haszpra, L., Barcza, Z., Bakwin, P. S., Berger,
B. W., Davis, K. J., Weidinger, T. 2001: Measuring
system for the long-term monitoring of biosphere/atmosphere
exchange of carbon dioxide. Journal of Geophysical
Research. Vol. 106D, 3057-3070.
Haszpra,
L., 1999: On the representativeness of carbon
dioxide measurements, Journal of Geophysical
Research, Vol. 104D, 26953-26960.
Barcza,
Z., 2001: Long term atmosphere/biosphere exchange
of CO2 in Hungary. Ph.D. Dissertation,
Eötvös Loránd University, Department
of Meteorology, Budapest, 2001.
|