next up previous contents
Next: Calibration Up: The direct flux measuring Previous: Introduction   Contents


The monitoring system

Measurements of the vertical flux of CO\( _{2}\protect \), H\( _{2}\protect \)O, momentum and sensible heat by eddy covariance at 82 m above the ground was facilitated by mounting an ultrasonic anemometer (Gill Solent Enhanced), an aspirated thermocouple (50 cm from the anemometer) for fast-response temperature measurements, and an additional air sampling tube for CO\( _{2}\protect \) and H\( _{2}\protect \)O measurements to the 4.4 m long instrument arm. The measuring system was developed and operated by László Haszpra (Hungarian Meteorological Service).

Measurements of CO\( _{2}\protect \) and H\( _{2}\protect \)O are made at \( \sim \)4 Hz using a fast response IRGA (LI-COR model LI-6262). Air is pumped through the 115 m long sampling tube and the analyzer at about 15 l min\( ^{-1} \), producing a pressure drop of approximately 45 kPa. Pressure fluctuations generated by the pump are damped by means of a 6 l buffer volume. Immediately behind the sample cell of the analyzer we measure pressure (MKS Instruments model 122A barotron), temperature and relative humidity (Vaisala HMD20YB). The pressure and temperature data are used to correct the instrument response for variations in these parameters (see below). The humidity data are used to determine the calibration function for water vapor measurements by the LI-COR 6262 IRGA.

The CO\( _{2}\protect \) analyzer runs in relative mode. Dry, synthetic air with a CO\( _{2}\protect \) mixing ratio of 330-340 ppm is used as a reference gas (Messer Hungarogáz) during the measurements. The flow rate is 5-10 cm\( ^{3} \) min\( ^{-1} \) through the reference cell of the IRGA. The analog output of the analyzer for CO\( _{2}\protect \) and H\( _{2}\protect \)O, as well as the signals of the pressure and temperature/humidity sensors are digitalized by the common A/D-RS232 converter.

A separate data acquisition computer (486 PC with a 40 MHz CPU, 4 MByte RAM, 1 GByte HD) is used to read data from the fast-response instruments (sonic anemometer, thermocouple, and IRGA). The computer communicates with the instruments through two standard serial ports: COM1 receives the data from the sonic anemometer while the two A/D-RS232 devices are controlled via COM2. The data acquisition cycle is triggered by the signal from the sonic anemometer. After reception of a data package (horizontal and vertical wind speed, wind direction and error code) the computer requests data from the aspirated thermocouple, the CO\( _{2}\protect \)/H\( _{2}\protect \)O analyzer and its accessory sensors (pressure, temperature/humidity), and the CO\( _{2}\protect \) profile analyzer. The position of the multiport valve of the profile system is also determined, which allows synchronization between the two independent data acquisition computers. The data acquisition software is written in Turbo Pascal language and runs under DOS 5.0. The eddy covariance system produces data at a rate of about 600 MByte/month, and the data are stored on a CD-R without compression.

Data coverage from the beginning of the EC measurement (end of April, 1997) until the failure of the IRGA (which caused data loss from the beginnig of 2000) was around 85%. After data screening it reduced to 78%. The average data coverage for the FLUXnet community is 69% (Falge et al., 2001).


next up previous contents
Next: Calibration Up: The direct flux measuring Previous: Introduction   Contents
root 2001-06-16