The LI-COR LI-6262 is calibrated against two standard gases with known concentrations
(320 ppm and 420 ppm) once a day, short before midnight. This method is more
common in the field of long term ecosystem carbon dioxide exchange studies than
the one described in section . The advantage of using
calibration gases is the avoidance of the complex calibration procedure described
in section
. The disadvantage is the interruption of
the data collection during the calibration period. The analyzer runs in absolute
mode.
Calibration starts at 23h 46min UTC+1, and finishes at midnight. The programmable timer switches between the standard gases in every 2 minutes so that each first minute is used to flush the sampling cell, while the signal measured during the second minute is integrated to determine the response of the instrument, to track the zero and span drift (LI-COR, 1996).
Factory calibration (fifth order polynomial for CO and third order
polynomial for H
O) is applied to convert the raw voltage signal into
mole fraction (LI-COR, 1996). Dry air mixing ratio is calculated in the same
way as it is described in section
to avoid using the
Webb-correction (Webb et al., 1980). Span and zero drifts are taken into account
using a linear stretch of the time series based on the measured mixing ratios
of the calibration gases. The validity of the linear approximation for the instrument
in the range of interest is described in section
.