Measurements of the vertical profiles of CO mixing ratio, temperature,
humidity and wind began at the end of September 1994.
Mixing ratios of CO are measured at 10, 48, 82 and 115 m above the
ground (Fig.
). Air is pumped through 9.5 mm (outer) diameter tubes
to a CO
analyzer located in the TV transmitter building. A 47 mm
diameter particle filter is located at the inlet of each tube. The set-up for
CO
analysis is very similar to that described by Zhao et al. (1997a),
which was used for the measurements reported in Bakwin et al. (1995, 1998).
Diaphragm pumps are used to draw air continuously through each of the tubes
from the four monitoring levels at a flow rate of about 2 l min
.
After the pump, the air at 40 kPa overpressure enters a glass trap for liquid
water which is cooled in a regular household refrigerator, to dry the air to
a dew point of 3-4
C. Liquid water is forced out through an orifice
at the bottom of each trap.
The four inlet tubes and the standard gases are connected to a computer controlled,
16-position valve, that selects which monitoring level or standard gas is sampled
by the analyzer. The valve head is protected by 7 mm in-line filters. Ambient
air flows continuously through the multiport valve so that the system is constantly
flushed. The (expensive) standard gases are shut off when not in use by means
of computer-controlled solenoid valves. The air leaving the multiport valve
through its common outlet is further dried to a dew point of about -25C
by passage through a 182 cm long Nafion drier, so that the water vapor interference
and dilution effect are less than 0.1 ppm equivalent CO
(Zhao et
al., 1997a). The Nafion drier is purged in a counter-flow arrangement using
waste sample air that has been further dried by passage through anhydrous CaSO
.
Analysis for CO is carried out using an infrared gas analyzer (IRGA)
(LI-COR model LI-6251). A constant sample flow rate of 100 cm
min
is maintained by a mass flow controller. The reference cell of the CO
analyzer is continuously flushed at a flow rate of 5-10 cm
min
with a compressed reference gas of 330-340 ppm CO
in air (Messer
Hungarogáz). The analyzer is calibrated by four standards covering 330-420 ppm
CO
, that were prepared by NOAA/CMDL (Kitzis and Zhao, 1999).
The basic measuring cycle is two minutes, consisting of one minute flushing
and one minute signal integration. Each one minute average and standard deviation
is based on 6-7 measurements. The multiport valve steps through the four monitoring
levels in eight minutes. Every 32 minutes, after four 8 minutes measuring cycles,
the standard gas with the lowest CO mixing ratio is selected and
analyzed for 2 minutes, and we term this measurement a zero. After every sixth
cycle (every 202 minutes, that is 4x8 min sampling and 2 min zero five times,
and one 4x8 min sampling without the zero) a full four-point calibration is
carried out. The reference and sample cells of the CO
analyzer are
not pressure or temperature controlled. The zero measurements are used to account
for any short-term drift of the analyzer due to changes in ambient pressure
or temperature. A quadratic response function is fit to each set of calibration
gas measurements. The zero offset and response function are linearly interpolated
in time to obtain values appropriate to calculate CO
mixing ratio
from the instrument response.
The off-line postprocessing of the profile data consists of the calculation
of the response functions for the CO analyzer and the conversion
of the voltage data into physical units. If the change in the response function
causes more than 2 ppm change between two consecutive calibrations, the data
for the period is rejected. Such periods are rare, and almost always caused
by significant change in room temperature. The usual change of the response
function is below 0.3 ppm. It should be noted that the drift equally influences
all monitoring levels, therefore the relative mixing ratio profile is correct
even if the absolute accuracy is temporarily lower than usual. As this type
of error is random, the long-term accuracy of the values is close to that of
the standards (about 0.1 ppm, Zhao et al., 1997b).
For data acquisition and system control for the CO profiles and meteorological
data we use a 286 PC with 1 MByte RAM and a 40 MByte hard disk. The analog signals
of the CO
analyzer and mass flow controller are read by a multiplexer-A/D
converter (PCL-711B). The data acquisition and system control software is written
in Turbo Pascal (Borland) and runs under MS DOS 5.0. During the data integration
period the computer consecutively reads data from the meteorological sensors
at each monitoring level through its serial port (RS232), then it reads the
CO
analyzer and the mass flow controller through the PCL-711B card.
The profile system generates 5 MByte/month of data which is stored on a floppy
diskette after compression. The data are mailed or carried to our laboratory
in Budapest, where data corrections and data processing are performed.